Constant temperature water bath with exocyclic
Thermo supplier

EZ-101
Product Code:0043201-000

External circulation while using water bath is available. Addressed to high temperature, silicon oil can be used as well.



Image of combination of EZ-101 and Immersion cooler

Main Application

Circulation to device which need temperature regulation at high temperature
Quality test at high temperature using water bath

Features

EZ-101 for high temperature range

Heat insulated water bath has small heat loss and can be used stably at high temperature. Cooling pipe immersion part is set and use of Immersion cooler enables setting at below 0ºC without wasting water bath space.

External circulation with variable flow rate, watering surface is made of all stainless steel

It is optimum to circulate into precise device by variable flow rate. Since all watering surface is made of stainless steel, heat resistance and sanitary quality is improved.


Specifications

ModelEZ-101
Temperature Range +60ºC - +200 ºC (*)
Temp. Control Accuracy ±0.1ºC - 0.3 ºC
Pump ability(50/60kHz) Max. Flow : 8/10L/min (variable)
Max. lifting height : 4/5m
Heater output 1200W
Thermoregulation method Pt. PID regulation (with auto tuning)
Minimum temperature setting unit 0.1ºC
Safety device / function Short/ overcurrent breaker, fuse (3A), floating type boil without water protection (with protection of chattering), thermosense type boil without water protection, thermal switch for motor protection, high and low temperature protection of sample, freezer protection (approx. +40ºC, only for EZL-81F), diagnosis of temperature sensor error (disconnection/ short), diagnosis of low water level error, alarm for high and low temperature protection of sample, mixing inside water bath
Dimensions inside bath 170 x 300 x 145Hmm
Water bath volume (water level 80%) Approx. 6L
Effective Dimensions inside bath 150 x 100 x 140Hmm
Dimensions 230 x 368 x 407Hmm
Weight Approx. 15kg
Power Supply AC100V 50/60Hz 13A
(*1) Temperature setting is available between -30ºC - +200 ºC. When setting temperature lower than RT, please use Immersion cooler together. There is special opening for Immersion cooler.
When setting below +7ºC, please use antifreeze (see option)
When setting around RT - +60ºC, please circulate cooling water (tap water) in cooling pipe. Temperature rise by work heat from mixing pump need to be minimized at below +60ºC.
When setting above +70ºC, please use heating medium for high temperature (see option).
Please do not use pure water or distilled water as heating medium. These may cause breakdown of machine.

Dimensions

Options

Heating medium for low ...  Antifreeze  Show Brine Blue

Use when needed together with low temperature water bath unit or immersion cooler

Heating medium for high ...  Silicon oil MA-100

Kinetic viscosity 100mPa sec (cSt), 18kg, for use higher than +100 ºC

Heating medium for high ...  Silicon oil MA-50

Kinetic viscosity 50mPa sec (cSt), 18kg, for use higher than +50 ºC

Related Products

Constant temperature wa ...  Coolnit bath  EL-15F

External circulation while using water bath, variable flow rate. Remote control and t ...

Constant temperature wa ...  Coolnit bath  EL-8F

External circulation while using water bath, variable flow rate. Remote control and t ...

Constant temperature wa ...  Thermo supplier  EZL-81F

External circulation while using water bath is available. Addressed to high temperatu ...

Constant temperature wa ...  Thermo circulator  ZL-100

Slim type chiller for sealed circuit which doesn't occupy much space. Thermoregulatio ...


home products faq about_us inquiry