Product Info
Constant temperature water bath with exocyclic coolnit bath
Coolnit bath EL-15F
Code No.: 0013967-000
External circulation while using water bath and variable flow rate. Can also be used for calibrating the thermometer. Remote control and temperature programming are available as an option.


Applications
●Temperature regulation of spectrophotometer (thermal block), etc.●Temperature regulation of culturing apparatus and analytical device
●Calibration of thermometer in water bath, Keeping the temperature of the sample
Down Load(Useful Information)
General CatalogFeatures
●Water bath is that free of projections and easy to place test tube stands, etc.●External circulation function by variable flow rate
●Remote control setting input is available as an option
Specifications
Model | EL-15F |
---|---|
Temperature range | -10℃ to +70℃ (*1) |
Control accuracy | ±0.1 to 0.3℃ (*2) |
Cooling capacity | approx. 290W (*3) |
Compressor(air cooled-type)output, Refrigerant | 150W, R134a |
Pumping capacity(50/60Hz) | Max.flow rate: approx. 8.2/10 L/min, Max. lift height: |
Heater | 800W |
Nozzle diameter | Outer diameter φ10mm |
Safety device/function | Short/Over current breaker, Floating type boil without water protection, High temperature |
Other Functions | Remote temperature setting input (DC 0 to 5 V), agitating inside the water bath |
Dimensions inside the bath | 250×395×150H mm |
Effective dimensions inside the bath | 250×250×150H mm |
Water bath capacity(water level 80%) | approx. 12L |
Dimensions | 310×560×575H mm |
Weight | approx. 35kg |
Power Supply | AC100V/14A (Need a step-down transformer) |
Standard Accessory | Stainless steel lid × 1 pc |
(*1)Please use heating medium for low temperature when setting at below 7℃. When using the optional program unit, temperature range is -20℃ to +80℃ (There may be a case of diffi cult to use at below -10℃ depending on operating condition). | |
(*2)Performance may not be maintained due to environmental temperature, power supply, etc. | |
(*3)When circulating fluid is +10℃, 50 Hz. | |
•Please do not use pure water or distilled water. These may cause breakdown of the machine. |
Dimensions

Accessories

| Circulation nozzle ex.dia. φ13mm
For changing nozzle diameter■Applicable Models:EL-8F, EL-15F ■:

| Caster
Set of 4, For EL-8F/15F CP-80R/150R CL-80R/150R■Qty:Set of 4 ■Applicable Models:EL-8F/15F, CP-80R ...

Antifreeze | Show Brine Blue
Use when needed together with the low temperature water bath unit or immersion cooler, 20 kg, 1 can, use when setting below +7℃■Weight:20kg, 1 can ■:
Related Products

Constant temperature water bath with exocyclic coolnit bath | Coolnit bath | EL-8F
【discontinued】External circulation while using water bath and variable flow rate. Can also be used for calibrating the thermometer. Remote control and temperature programming is available as an option.■Temperature range:-10℃ to +70℃ (*1) ■Control accuracy:±0.1 to 0.3℃ (*2)

Constant temperature water bath with exocyclic coolnit bath | Thermo supplier | EZ-101
External circulation while using the water bath is available. Addressed to high temperature. Silicon oil can be used as well. EZ-101 for high temperature range.■Temperature range:+60℃ to +200℃ (*1) (*2) ■Control accuracy:±0.1 to 0.3℃ (*3)

Constant temperature water bath with exocyclic coolnit bath | Thermo supplier | EZL-81F
External circulation while using the water bath is available. Addressed to high temperature. Silicon oil can be used as well. EZL-81F for low to high temperature range.■Temperature range:-10℃ to +100℃ (*1) (*2) ■Control accuracy:±0.1 to 0.3℃ (*3)