e-TAITEC.com

Product Info

Constant temperature water bath with exocyclic coolnit bath

Thermo supplier EZL-81F

Code No.: 0043200-000

External circulation while using the water bath is available. Addressed to high temperature. Silicon oil can be used as well. EZ-101 for high temperature range and EZL-81F for low to high temperature range.


Applications

Circulation to device which needs temperature regulation at high temperature.
Quality test at high temperature using the water bath.
Moisturizing and Cooling for samples.

Down Load(Useful Information)

TAITEC general catalog-Thermo supplier EZ-101/EZL-81F (PDF)

Features

●Silicon oil can be used in the heat insulated water bath
●Temperature in low temperature range can be adjusted in EZL-81F
●External circulation function by variable flow rate, Wetted parts are made of stainless steel

Specifications

ModelEZL-81F
Temperature range-10℃ to +100℃ (*1) (*2)
Control accuracy±0.1 to 0.3℃ (*3)
Cooling capacityapprox. 140W (*4)
Compressor (air cooled-type)output, Refrigerant80W, R134a
Pumping capacity (50/60Hz)Max.flow rate: approx. 8/10 L/min (variable)
Max. lift height: approx. 4/5 m
Heater800W
Nozzle diameterOuter diameter φ10mm
Safety device/functionShort/Over current breaker, Fuse (3 A), Floating type boil without water protection (with protection of chattering), Thermosense type boil without water protection, Thermal switch for motor protection, High and Low temperature protection of sample, Freezer protection (approx. +40℃), Diagnosis of temperature sensor error (disconnection/short), Diagnosis of low water level error, Alarm for high and low temperature protection of sample
Other FunctionsAgitating inside water bath, Cooling pipe (Outer diameter φ8mm) (*2)
Dimensions inside bath210×300×145H mm
Effective dimensions inside bath150×100×140H mm
Water bath capacity (water level 80%)approx. 7.3L
Dimensions270×368×617H mm
Weightapprox. 35kg
Power SupplyAC100V・12A
Standard AccessoryStainless steel lid×1
(*1)When setting below +7ºC, please use antifreeze. When setting above +70ºC, please use heating medium for high temperature.
(*2)When setting around +40ºC to +60ºC, please circulate cooling water (tap water) in cooling pipe. Temperature rise by work heat from mixing pump need to be minimized at below +60ºC.
(*3)Performance may not be maintained due to environmental temperature, power supply, etc.
•Please do not use pure water or distilled water. These may cause breakdown of machine.
(*4)When environmental temperature is +25ºC and circulating fluid is 10ºC, 50Hz.

Dimensions

Accessories

Silicone oil | MA-50
Kinetic viscosity 50 mPa sec (cSt), 18 kg, for precise temperature control and use higher than +70℃.
■: ■:
Silicone oil | MA-100
Kinetic viscosity 100 mPa sec (cSt), 18 kg, for use higher than +70℃
■: ■:
Antifreeze | Show Brine Blue
Use when needed together with the low temperature water bath unit or immersion cooler, 20 kg, 1 can, use when setting below +7℃
■Weight:20kg, 1 can ■:

Related Products

Constant temperature water bath with exocyclic coolnit bath | Coolnit bath | EL-8F
External circulation while using water bath and variable flow rate. Can also be used for calibrating the thermometer. Remote control and temperature programming is available as an option.
■Temperature range:-10℃ to +70℃ (*1) ■Control accuracy:±0.1 to 0.3℃ (*2)
Constant temperature water bath with exocyclic coolnit bath | Coolnit bath | EL-15F
External circulation while using water bath and variable flow rate. Can also be used for calibrating the thermometer. Remote control and temperature programming is available as an option.
■Temperature range:-10℃ to +70℃ (*1) ■Control accuracy:±0.1 to 0.3℃ (*2)
Constant temperature water bath with exocyclic coolnit bath | Thermo supplier | EZ-101
External circulation while using the water bath is available. Addressed to high temperature. Silicon oil can be used as well. EZ-101 for high temperature range and EZL-81F for low to high temperature range.
■Temperature range:+60℃ to +200℃ (*1) (*2) ■Control accuracy:±0.1 to 0.3℃ (*3)
Translate »