Product Info
Active gas ventilation type large CO2 incubator for stack plate
Prescyto MG-1300M-2A
Code No.:
CO2 incubator for large scale cell culture and production of vaccines, monoclonal antibodies, etc. by stack plates

Applications
●For large static scale cell culture●For production of vaccines
●For production of monoclonal antibodies
Features
High capacity for static cell-culture
MG-1300M-2A is an incubator optimized for 10 or 40 layered stack plates: ThermoFisher [Nunc] Cell Factory and Corning CellSTACK.Precise gas monitoring and regulation
IR CO2 sensor and zirconia O2 sensor installed in MG-1300M-2A are reliable for constant cell culture with long-term stability.Temperature control optimized for culture with stack plates
A cross-flow fan in chamber and heating system for gas flow-line promote ideal culture condition in stack plates.Built-in PC for validation documentation
Operating parameters (temperature, CO2 concentration and gas flow) are graphically displayed on the panel and recorded on the built-in PC based on GMP standards.Customizations
We can make customized models upon user’s request: such as 4 ch model, O2 contorol model, no-heater model for incubation room etc. Please contact us.TAITEC MG-1300M and CBR-500S DemoMovie
Specifications
Model | MG-1300M-2A |
---|---|
Temp.control range | Ambient +5℃ to 40℃ |
Temp. Control Accuracy | ±0.5℃(*1) |
Dimensions (interior) | 1100×740×1300Hmm |
Dimensions | 1280×950×1930Hmm |
Ambient temp. range | +5℃ to +30℃ |
Heating method | Air jacket. 4 heaters on each side in chamber. Heating gas inflow line. |
Flow control | Mass flow controller for each channel |
CO2 Sensor/Control range | IR sensor, 3.0 to 8.0% (Increment: 0.1%) |
O2 Sensor/Control range | Zirconia sensor, 1 to 25% |
Gas supply | Max. 2L×2 lines (Compatible with 40 layers x 2 lines), Min. 0.1L setting unit: 0.02L in use |
Safety devices | Control PCB alarm (Buzzer), Heater cut off when the door open |
Safety function | High/low temperature, Pressure of gas and air, CO2 concentration, O2 concentration, Flow rate for each channel, Door open/close with buzzer |
Material of chamber | stainless (sus304) |
Power supply | AC220V/4A (*2) |
Record contents | Log of time, Temperature, CO2 concentration, Flow rate |
Record | Recording at built -in PC (*3), Output at USB memory. |
Graphical display | Recorder contents |
Required gus supply | CO2 gas line |
(*1)This value is under the condition of room temp. at +25℃ and, setteing temp. at +37℃ and 5% of CO2 concentration. | |
(*2)The power supply of built-in PC should be supplied from another outlet. | |
(*3)Not complied with Part11. | |
●This product is build-to-order manufacturing. |
Related Products

CO2 incubator | Prescyto | MG-71C
Multi-gas incubator equipped with a new "Low oxygen booster" function that makes Hypoxia conditions return quickly after the door opens and closes (4-times faster than the conventional one. *Patent pending).■Temperature range/accuracy:5℃ above RT to 50℃, ±0.2℃ (*1) ■Heating method:Air Jacket type (6 Heaters)

Active-gas type CO2 incubator | Prescyto | MG-71C-A
Our proprietary technology "Active Gas Ventilation" CO2/Multi-gas incubator. Optimal for Large scale and Hypoxic culture with Multilayer culture plates■Temperature range/accuracy:5℃ above RT to 50℃, ±0.2℃ (*1) ■Heating method:Air Jacket type (6 Heaters)

Active-gas type multi-gas incubator | Prescyto | MG-71M-A
Our proprietary technology "Active Gas Ventilation" CO2/Multi-gas incubator. Optimal for Large scale and Hypoxic culture with Multilayer culture plates■Temperature range/accuracy:5℃ above RT to 50℃, ±0.2℃ (*1) ■Heating method:Air Jacket type (6 Heaters)

| Manual manipulator with carts | HM-40
【discontinued】Option of manual manipulator with carts to handle 40 layer stack plate in filling and emptying of culture broth and / or trypsin treatment.■: ■: